5MP DOME E5 LCD for Mammography
The Dome E5 grayscale display is intended for the most demanding diagnostic applications, including mammography and radiology, in both portrait and landscape modes. The Dome E5 display provides seamless image manipulation and smooth window leveling, and supports open architecture graphics standards for three-dimensional and volumetric rendering applications. With full DICOM conformance, automatic calibration, and remote monitoring and maintenance, the Dome E5 is ideal for the most demanding PACS.
Discontinued Product

MASTER DISTRIBUTOR

24X7 SUPPORT

COMPETITIVE PRICING

SAME DAY SHIPPING

VALUABLE CONSULTATION
Product Information
![]() |
Dome E5 Cut Sheet |
![]() |
Dome Displays Technical Specs |
![]() |
Dome Video Board Compatibility Matrix |
![]() |
Dome Technology Comparison |
Case Studies
![]() |
Nebraska Medical Chooses Dome |
![]() |
Mission Health Chooses Dome |
![]() |
St. Lukes Health System Chooses Dome |
![]() |
Kaleida Health Chooses Dome |
White Papers
![]() |
ACR Guidelines |
![]() |
Long Term Reliability |
![]() |
Monitor Math |
Doing It Right
![]() |
Doing it Right S10 |
![]() |
Doing it Right Calibration |
![]() |
Doing it Right Front Sensors |
![]() |
Doing it Right Clear Image |
![]() |
Doing it Right Monitor Specs |
![]() |
Doing it Right Precision LUTs |
Calibration kit
Calibration Kit – DOME Photometer . American College of Radiology guidelines still recommends the periodic use of a calibration kit to insure that diagnostic integrity is maintained over the life of the monitor. Some states now require the use and documentation of a calibration kit. Do not confuse this device with the front sensors on most diagnostic medical displays. They are not the same.
What You Get:
NDS DOME E5 5MP Monochrome LCD Monitor Only, No Video Card. E5 -2048×2560 landscape/portrait – DVI input, CXtra software, display driver and cables included. 5 year Manufacturer’s Warranty. Premium Service plans are available. Part Number – 997-5715-00-1NN.
List Price – $12,795.00
Specification
LCD Characteristics | |
---|---|
Technology | TFT AMLCD Dual Domain IPS, 3 subpixels per pixel |
Screen Size (diagonal) | 21.3″ (541mm) |
Display Resolution (pixels) | 2048 x 2560 Portrait, 2560 x 2048 Landscape |
Pixel Pitch | 165μm, 154 dpi |
Viewing Angle | 170° horizontal and vertical |
Response Time | 50 ms typical |
Optical Characteristics | |
Brightness | 1000 cd/m² typical |
Default Calibrated | 500 cd/m² |
Contrast Ratio | 800:1 typical |
DICOM LUT | 3061 unique shades of gray |
DICOM Characterization | LCD panels individually factory characterized with a NIST-traceable photometer. Characteristic data is stored in the display and remains available throughout the product life. |
DICOM Calibration | DICOM calibration performed by Dome CXtra™ using the RightLight™ internal photometer. DICOM calibration LUT stored in the display and automatically maintained for the life of the display. |
DICOM Conformance Testing | DICOM conformance maintained by RightLight™ internal photometer. Independent conformance testing using Dome CXtra™ with an external photometer. |
Panel Connectivity | |
Video Input | Dual Link DVI-D |
Pixel Data Sizes | 8-bit and 10-bit per pixel grayscale, 24-bit color mapped to luminance |
VGA Compatibility | 640 x 480 to 1280 x 1024 |
USB Interface | Built-in USB 2.0 HUB on stand |
Physical Characteristics | |
Display Size without stand
(H x W x D) |
18.5″ x 15.2″ x 4.1″
(470mm x 387mm x 103mm) |
Weight without stand | 14 lbs (6.4 kg) |
Weight with stand | 22 lbs (10.0 kg) |
Mounting Options | 100 x 100 mm VESA mounting standard. Desktop stand includes height, pivot, swivel, and tilt adjustment. |
Power Input | 100 to 240 VAC 50-60 Hz |
Power Consumption | 55 Watts typical, 72 Watts max |
Operating Range | 0 to 40 degrees C |
Certifications | |
Medical | FDA 510(k) |
Safety & Emissions | CSAus, EN60601-1, FCC, CE, CCC, TUV-GM, VCCI, KCC |
Environmental | RoHS, WEEE, REACH |
Service and Support | |
Warranty | 5-year standard warranty |
Backlight Warranty | 10-year standard warranty |